v 1.04
Clear Cal Report
This button clears the current calibration report.
Note: This report should be cleared prior to starting a calibration session and should be read (downloaded and archived immediately following the calibration session. This report is not saved within the Multi-Tracer system.
Module Voltage, Current, & Aux Input
These buttons open the respective calibration dialog for each type of measurement.
Calibration Dialog
Use the channel selection drop down to select the specific channel currently being calibrated.
The "Actual" input box contains the actual value of the signal. This value is manually entered and must come from an independent source such as calibration reference standards.
The "Measured" values is the real-time value of this signal as measured by the Multi-Tracer.
The "Load On" check box turns the module channel fully on to a near short-circuit condition. Only used for module current calibration.
Adjust Button
This button adjusts the calibration constant so that the measured value matches the given actual value. An entry is also made in the calibration report.
Record Button
The record button only reads and records an entry in the calibration report. This measurement using the existing calibration constant. This is useful for recording calibration values at multiple points for documentation purposes. It has no effect on the behavior of the Multi-Tracer system.
Next Button
Qucikly advances to next channel.
Save Button
Saves the adjustments of all channels.
Cancel Button
Deletes the adjustments of all channels and restores the orignal calibration (as if the dialog had not been used).
Done Button
Closes the dialog.